Xanthine,
C
5H
4N
4O
2
152.11White,crystalline powder.Decomposes on heating.Slightly soluble in water and in alcohol;soluble in sodium hydroxide TS;sparingly soluble in diluted hydrochloric acid.When subjected to the murexide reaction,a purple color is produced with the ammonia,but on the subsequent addition of fixed alkali hydroxides,the color is not discharged but is changed to violet.
Residue on ignition (Reagent test):
negligible,from 100mg.
Xanthydrol,
C
13H
10O
2
198.22Pale yellow,crystalline powder.Insoluble in water;soluble in alcohol,in chloroform,and in ether.Soluble in glacial acetic acid,forming a practically colorless solution;but when the powder is treated with diluted hydrochloric acid,a lemon-yellow color is produced.
Residue on ignition
Ignite 500mg with 0.5mLof sulfuric acid:the residue weighs not more than 10mg (2.0%).
Xylene,
C8H10106.17Use ACSreagent grade.
m-Xylene,
C6H4(CH3)2106.17[108-38-3]Clear,colorless,flammable liquid.Insoluble in water;miscible with alcohol and with ether.Use a suitable grade.
o-Xylene,
C
8H
10
106.17Clear,colorless,mobile,flammable liquid.Insoluble in water;miscible with alcohol and with ether.
Assay
When examined by gas-liquid chromatography,it shows a purity of not less than 95%.The following conditions have been found suitable for assaying the substance:A3-mm ×1.8-m stainless steel column packed with 1.75%hydrated aluminum silicate plus 5.0%diisodecyl phthalate on support S1.Helium,having a flow rate of about 27.5mLper minute is the carrier gas,the detector temperature is about 280

,the injection port temperature is about 180

,and the column temperature is 80

.Aflame-ionization detector is employed.
p-Xylene,
C
8H
10
106.17Colorless liquid.
Assay
Inject an appropriate volume into a gas chromatograph (see
Chromatography á621ñ)equipped with a flame-ionization detector,helium being used as the carrier gas.The following conditions have been found suitable:a 0.25-mm ×30-m capillary column coated with a 1-µm layer of phase G14;the injection port temperature is maintained at 130

;the detector temperature is maintained at 300

;and the column temperature is maintained at 50

and programmed to rise 10

per minute to 100

.The area of the C
8H
10peak is not less than 99%of the total peak area.
Xylene Cyanole FF,
C
25H
27N
2NaO
6S
2
538.61Gray-blue to dark blue powder.Soluble in water.
Assay
Transfer about 50mg,accurately weighed,to a 100-mLvolumetric flask,dissolve in water,dilute with water to volume,and mix.Transfer 2.0mLof the solution to a 50-mLvolumetric flask,dilute with pH7.0phosphate buffer to volume (See
Solutions,in this section),and mix.Using a suitable spectrophotometer,1-cm cells,and water as the blank,record the absorbance of the solution at the wavelength of maximum absorbance at about 614nm.From the observed absorbance,calculate the absorptivity (see
Spectrophotometry and Light-scattering á851ñ):the absorptivity is not less than 55.9,corresponding to about 83%of C
25H
27N
2NaO
6S
2.
Xylose,
C5H10O5150.13Use a suitable grade.