Chromatographic system(see Chromatography á621ñ)
The gas chromatograph is equipped with an on-column,temperature-programmable injector,a flame-ionization detector maintained at about 275

,and a 0.32-mm ×30-m column bonded with a 0.5-µm layer of phase G42.The column temperature is programmed to be maintained at about 80

for 0.5minute,then to increase to about 220

at a rate of 20

per minute,and to hold at about 220

for 10minutes.The injection port temperature is programmed to be maintained at about 85

for 0.5minute,then to increase to about 225

at a rate of 20

per minute,and to hold at about 225

for 10minutes.Helium is used as the carrier gas at a flow rate of about 2.3mLper minute.Chromatograph the
System suitability solution,and record the peak responses as directed for
Procedure:the relative retention times are about 0.9for triethyl citrate and 1.0for acetyltriethyl citrate;the resolution,
R,between triethyl citrate and acetyltriethyl citrate is not less than 1.5;and the relative standard deviation for replicate injections is not more than 2.0%determined from both the triethyl citrate and acetyltriethyl citrate peaks,based on area percent calculation.
Procedure
[NOTEUse peak areas where peak responses are indicated.
]Inject 1µLof the
Assay preparationinto the chromatograph,record the chromatogram,and measure all of the peak areas,excluding the solvent peak.Calculate the percentage of C
12H
20O
7in the portion of Triethyl Citrate taken by the formula:
100(A/B),
in which
Ais the triethyl citrate peak response;and
Bis the sum of the responses of all the peaks.